Analisis Ergonomi Dan Keamanan Tempat Duduk Penggunaan Crane Pengangkat Elektrik Dalam Proses Transfer Pasien Penyandang Disabilitas

Ayu Setyaning Sayekti Poesoko, Bambang Setyono, Fachrul Ramadhan Chalifatullah

Abstract

ABSTRACT

Disability is a condition in which an individual experiences limitations or an inability to perform activities in ways that are considered normal by most people. These limitations are usually caused by impairments or a decline in the individual's abilities. As a result, the person may require assistance from others to manage daily life. This study aims to analyze ergonomic and safety aspects of the use of an electric lifting crane seat in the process of transferring patients with disabilities. The research employs the RULA method with weight variations of 40 kg, 60 kg, and 80 kg. The results indicate that the electric lifting crane in transferring patients weighing 40-60 kg received a score of 2, indicating minimal risk and no immediate need for improvement. Meanwhile, for patients weighing 80 kg, the device scored 3, indicating a low risk in the near future. The transfer belt pad for the wheelchair on the electric lifting crane is deemed appropriate according to the SNI standard (09-4663-1998), as no physical defects were found.

 

 

Keywords

Kata kunci: electric lifting crane, RULA, patient transfer, score, Ergonomics

Full Text:

PDF

References

R. Pasciana, “Pelayanan Publik Inovatif Bagi Penyandang Disabilitas: (Studi Kasus di Kabupaten Garut),” Sawala J. Adm. Negara, vol. 8, no. 2, pp. 192–203, Dec. 2020, doi: 10.30656/sawala.v8i2.2857.

K. Kjellberg, M. Lagerström, and M. Hagberg, “Patient safety and comfort during transfers in relation to nurses’ work technique,” J. Adv. Nurs., vol. 47, no. 3, pp. 251–259, Aug. 2004, doi: 10.1111/j.1365-2648.2004.03089.x.

I. M. L. Batan, “PENGEMBANGAN KURSI RODA SEBAGAI UPAYA PENINGKATAN RUANG GERAK PENDERITA CACAT KAKI,” J. Tek. Ind., vol. 8, no. 2, pp. 97–105, Feb. 2007, doi: 10.9744/jti.8.2.97-105.

S. Koyama et al., “Comparison of two methods of bed-to/from-wheelchair transfer in patients with hemiparetic stroke,” 2020, Fujita Medical Society: 3. doi: 10.20407/fmj.2019- 016.

W. Yudiantyo, R. Wawolumaja, and S. Soly, “Perancangan Fasilitas Penunjang untuk Pemindahan Pasien dari/ke Kursi Roda ke/dari Tempat Tidur Melalui Pendekatan Ergonomi,” J. Integr. Syst., vol. 6, no. 2, pp. 210–225, Dec. 2023, doi: 10.28932/jis.v6i2.7554.

Y. Alexander, “Perancangan Alat Bantu Perpindahan Barang yang Ergonomis (Studi Kasus Di PT.”X”, Bandung),” vol. 1, no. 2.

J. Setiawan, G. B. Mandegani, and E. Y. Rufaida, “Analisis Kesesuaian Kursi Pembatik Terhadap Kondisi Antropometri Pekerja Batik Tulis,” Din. Kerajinan Dan Batik Maj. Ilm., vol. 31, no. 2, p. 113, Apr. 2016, doi: 10.22322/dkb.v31i2.1077.

L. Dewiyani and M. K. -, “PERANCANGAN TATA LETAK DAN FASILITAS HUNIAN RUKOST BERDASARKAN KEBUTUHAN KONSUMEN UNTUK MENINGKATKAN PANGSA PASAR DENGAN MENGGUNAKAN ASPEK ERGONOMI,” SPEKTRUM Ind., vol. 16, no. 1, p. 77, Apr. 2018, doi: 10.12928/si.v16i1.9782.

A. Fajariansyah, “RANCANG BANGUN DAN INOVASI ELEKTRIK CRANE GAYA ANGKAT MAKS 500 KG. (PERAWATAN),” 2020.

P. Indonesia, “Permenkes Nomor 4 Tahun 2014.pdf.” Menteri Kesehatan Republik Indonesia, 2014.

D. M. Pangaribuan, “Analisa Postur Kerja Dengan Metode RULA Pada Pegawai Bagian Pelayanan Perpustakaan USU Medan.” 2010.

Refbacks

  • There are currently no refbacks.